Checking Work before Glove Box Operation
1. Check if there is nitrogen in the
nitrogen bottle. The nitrogen should be replaced in time. The output pressure
of nitrogen is set to 0.5mpa.
2. Check whether the vacuum pump is
normally turned on.
3. Check whether the circulation inside the
glove box is open. (Open one of Pure column - I or purification column - II)
4. Check that the upper limit pressure
should be +1 and the lower limit pressure is 0.
5. Check whether the oxygen and dew point
are normal (oxygen is below 0.1ppm, dew point is above -40)
6. The vacuum pump of antechamber should be in the normally open state.